The HI9811-51 portable meter features a large LCD which displays either pH, EC, TDS or temperature readings along with tutorial instructions. The HI9811-5 is a simplistic pH/EC/TDS meter that is calibrated manually to a single point with the use of two trimmers. pH is calibrated to pH 7.01 while EC/TDS is calibrated to either 1.41 mS/cm (1413 µS/cm) or 1500 ppm. The LCD screen has battery life indicator as well as on-screen tutorial messages.
The supplied HI1285-51 probe is a polypropylene body, amplified pH electrode with a built-in EC/TDS and temperature sensors. The amplifier for the pH electrode prevents interference from humidity and electrical noise from common sources including from motors, ballasts or pumps. The HI9811-5 is a versatile meter best suited for agricultural, greenhouse and hydroponic applications.
HI1285-51 polypropylene body amplified pH electrode with built in EC/TDS and temperature sensors
The HI9811-51 is supplied with a polypropylene body pH/EC/TDS/temperature probe. The pH, EC, TDS, and temperature sensors are housed in a single body that connects to the meter with a DIN connector. This probe is specially engineered with a fiber junction and gel electrolyte to produce a fast response with reduced contamination.
Amplified pH electrode
The pH electrode circuit has a built in amplifier that will reduce the effects of electrical noise on the high impedance pH measurement. Examples of sources of electrical noise include motors, ballasts, and pumps which are common in greenhouses.
Amperometric EC/TDS sensor
The EC/TDS readings are performed by an amperometric sensor. An alternating voltage is applied to the sensor and the amount of current that passes between the two stainless steel pins is dependent upon the amount of salts (fertilizer) present. A greater amount of salt present results in an increase in conductance.
Polypropylene body
The polypropylene body houses all the sensors in a single body design and is durable. The probe is gel filled for maintenance free operation. It does not have to be refilled periodically.
- 3 sensors in a single probe
- Pre-amplified pH electrode for resistance to electrical noise
- Gel filled maintenance free pH electrode
Features at-a-glance
- Water Resistant
- Automatic Temperature Compensation
- All readings are compensated for variations in temperature
- Low Battery Indicator
SKU | HI9811-51 |
---|---|
Product Name | pH/EC/TDS/Temperature meter with HI1285-51 electrode; EC range: 0 to 6000 µS/cm, TDS range: 0 to 3000 ppm (mg/L) |
Quote Required | Yes |
pH Range | 0 to 14 pH |
pH Resolution | 0.1 pH |
pH Accuracy | ±0.1 pH |
pH Calibration | manual, 1-point through offset trimmer |
EC Range | 0.00 to 6000 µS/cm |
EC Resolution | 10 µS/cm |
EC Accuracy | ±2% F.S. |
EC Calibration | manual, 1-point through slope trimmer |
TDS Range | 0 to 3000 ppm (mg/L) |
TDS Resolution | 10 ppm (mg/L) |
TDS Accuracy | ±2% F.S. |
TDS Calibration | manual, 1-point through slope trimmer |
EC/TDS Temperature Compensation | automatic from 0 to 50°C (32 to 122°F) |
Temperature Correction Coefficient | β = 2% /°C |
EC to TDS Conversion Factor | 0.5 ppm (mg/L) = 1 µS/cm |
Temperature Range | 0.0 to 70.0°C |
Temperature Resolution | 0.1°C |
Temperature Accuracy | ±0.5°C |
Multi-parameter Probe | HI1285-5 combination, amplified pH/EC/TDS/temperature probe with built-in temperature sensor, 8-pin DIN connector and 1 m (3.3′) cable |
Battery Type/Life | 9V / approximately 150 hours of continuous use |
Environment | 0 to 50°C (32 to 122°F); RH max 100% |
Dimensions | 145 x 80 x 36 mm (5.7 x 3.1 x 1.4”) |
Weight | 230 g (8.1 oz.) |
Ordering Information | HI9811-5 is supplied with HI1285-5 multi-parameter probe, HI70007 pH 7.01 buffer solution sachet, HI70032 1382 ppm calibration solution sachet, HI70031 1413 µS/cm calibration solution sachet, HI700661 electrode cleaning solution sachets (2), 9V battery, instructions and rugged carrying case. |
Use with Hanna products with best results:
Cleaning: Hanna Cleaning Solution General Purpose 500ml
Storing: Hanna Electrode storage solution 500ml
Buffering EC: Hanna 1413 µS/cm EC Solution 500ml
Buffering pH:
Hanna pH 4.01 Calibration Buffer 500ml
Hanna pH 7.01 Calibration Buffer 500ml
Hanna pH 10.01 Calibration Buffer 500ml
* Taking proper care and following instructions extends the lifetime of your meter and probe.